The Determination of the Total Efficiency for NaI(Tl) Detector by GATE Simulation
View/ Open
Date
2019Author
YAVUZKANAT, Nuray
GÜNGÜR, Didem
YALÇIN, Sezai
Metadata
Show full item recordAbstract
In this study the total efficiency of the NaI(Tl) detector have been calculated by using Geant4 based GATE simulation program. The simulation was performed using point and disc isotropic gamma-ray sources at various distance between source and detector in the photon energy range 50-3000 keV. Results were compared with different analytical and Monte Carlo calculations obtained in the time of the periods 1958-2018. The calculated and simulated data indicated good agreement with finding more sensitive GATE simulation result in the lower energy range from 50 keV to 150 keV.